Basic information
Main structure:consists of control part,motor,Contra Angle head,power adapter and charging base.
Application:Used in root canal therapy,root canal preparation stage shaping and cleaning,the product does not contain the root tip positioning function. Electrical safety classification:Class Il,internal power supply:
Has a type B application section
Product Code:XLY-Q200
Software release:LX037-GZ-101.1.0.0-170324
|
Code |
XLY-Q200 |
|
Power supply mode |
Rechargeable |
|
Functional characteristics |
Portable Digital display type Nine-memory program Quadruple inversion mode Speed optional Torque optional |
Technical parameters
Type B application section
Shell protection:IPX0
Classification according to the degree of safety for use with flammable anesthetic gases mixed with air and nitrous oxide:Non-APIAPG type.
*Body
Size:--------------------------Diameter,27×215mm
Classification:--------------------Common,Type B
Running mode:--------------------Continuous running Standard operating voltage:---------d.c.3.7V
*Power adapter
Input voltage: 100V-240V~50/60Hz
Output voltage:d.c.5V,1A
Specifications:GC09-050100-5A
*Battery
Type:Lithiumion battery Capacity:800mAh
Charging voltage:3.7V
Type specification:INR 14500
*Program parameter
Procedure: 9 kinds
Working Mode: 4 kinds Standard head:16:1
Net weight: 265g
Rotate speed: 140-550 r/min Torque:0.3-4.0N.cm
*Transportation and storage environment
a)Environment temperature range:-20℃~+55℃;
b)Relative humidity range:≤80%;
c)Atmospheric pressure range:70kPa~106kPa;
*Operating environment
a)Environment temperature range:5℃~40℃;
b)Relative humidity range:≤80%;
c)Atmospheric pressure range:70 kPa~106kPa;
d)power supply:d.c.3.7V(-10%~+5%)
e)Avoid using this product in an environment with electromagnetic interference(such as near a microwave oven)
f)The charger of this product shall meet the requirements of GB9706.1-2007.